Ⅰ.Confirmation of Equipment Design :
| 1 | Equipment | |
| 1.1 | Total volume
35L |
Use Fermentation tank Bioreactor |
| Working volume | 40-70% Quantity: 1 | |
| 1.2 | Liquid-adding container | Heat-resistant silica glass bottles
1L : 3pcs, 2L : 1pc, 10L 1pc, 20L : 1pc |
| 1.6 | Process piping | Enclosed installation
Open-shelves installation |
| 1.7 | Air pretreatment | Pressure-relief device No
Pre-filtration and pressure-relief devices No |
| 1.8 | Vapor pretreatment | Industrial steam filtration, hydrophobic and stress systems
No Pure steam filtration, hydrophobic and pressure relief systems required No |
| 1.9 | Cooling water pretreatment | First-level screen filter |
| 1.10 | Other | Air , Nitrogen, carbon dioxide, oxygen decompression device No |
| 1.11 | The buyer’s confirmation: | |
| 2 | Application | |
| 2.1 | Biological species | Bacteria(bacteria) Plant cells Mammlian Cells |
| 2.2 | Cultivating methods | Batch(Batch culture) Suspension culture
Microcarrier culture Fed batch(Batch supplement culture) |
| 2.3 | The buyer’s confirmation: | |
| 3 | Delivery date and installation site | |
| 3.1 | Equipment delivery date | 90-120 working days |
| 3.2 | The installation site | Customer will install the equipments by themselves |
| 3.3 | The buyer’s confirmation: | |
| 4 | Reactor/fermentation tank | |
| 4.1 | Tank | total volume: 35L Working volume Full volume
Working volume: 70% Diameter ratio: 1: 2.2-3 1: 1.5-1.8 ………. Design tank pressure: 0.3MPa …3…. barg Tank’s internal bile material: 316L 304 Test report of tank materials: 3(SGS):Top cover, straight tube body, lower head No Internal polishing accuracy: Ra0.2 Ra0.4 Ra0.8 Internal electropolishing Internal pickling passivation External surface polishing treatment: Ra0.8 sand blasting Internal polishing test report: 3(SGS):Top cover, straight tube body, lower head No Pressure vessel certificate: provide No Conform to GMP standard: Provide relevant documents No Viewing mirror: Side-view mirror Top-view mirror No floodlight: Side installation Top installation No Sterilization temperature: 100-130 °C,The sterilization time: 0-60min,The highest temperature: 135 °C Sterilization mode: In situ automatic sterilization (Can be sterilized separately for empty and full vessel) Manual operation Sampling valve: Denmarkkeofitt Bürkert No Sampling valve sterilization mode:Manual sterilization Tank bottom valve: Bürkert GEMÜ No Tank bottom valve sterilization mode:Automatic sterilization There are three optional filling liquid interfaces; A shallow aeration tube interface; A deep ventilation tube interface; A top flame inoculation interface;A side seed transfer interface Cleaning: high pressure cleaning interface is installed at the top; Tail gas condenser: yes No Exhaust gas incinerator: yes No |
| 4.2 | Jacket | Design pressure: 0.35Mpa
Jacket material: 304 stainless steel Temperature control jacket:Coverage ratio 80%,Internal optimization diversion structure Heat preservation jacket: yes No External surface treatment: buffing Ra0.8 matt |
| 4.3 | Top cover | Flat-topped tank cover, bolt – type compaction device;
Pneumatic open-cover mode Electrodynamic open-cover mode |
| 4.4 | The buyer’s confirmation: | |
| 5 | Mixing1 system | |
| 5.1 | Mixing type
Mechanical seal system |
The bottom double mechanical seal,Steam sterilization treatment
The top magnetic drive is not mechanical seal, the performance is better, the cost is high. |
| 5.2 | Mixing shaft type | Single axis, micro taper high speed type |
| 5.3 | Mixing speed | Speed range: 20-200rpm, stepless speed regulation |
| 5.4 | The mixing slurry | 2 grade elephant ear with 4-blade impeller or customer requirements |
| 5.5 | material | Internal tank and all contact with the material both are 316L |
| 5.6 | The buyer’s confirmation: | |
| 6 | Ventilation system | |
| 6.1 | Ventilation mode | bottom ventilation top ventilation The control interface can be selected in two modes
The bottom ring gas distributor The bottom l-shaped gas distributor |
| 6.2 | Filter | Air intake filter Exhaust filter
Automatic on-position sterilization Manual on-position sterilization on-position integrity testing off-position integrity testing None Brands:Sartorius Domnick Hunter |
| 6.3 | Gas species | air oxygen nitrogen carbon dioxide |
| 6.4 | Flow meter | Rotor flow meter + solenoid valve Gas mass flow meter + rotor flow meter
brands: Vogtlin Bürkert Cheng Feng |
| 6.5 | The buyer’s confirmation: | |
| 7 | Feeding System | |
| 7.1 | Feeding type | acid base antifoam Nutrient solution 1 + feeding material inoculation liquid
Note: usually, acid and alkali only use one peristaltic pump, the interface is optional, so the actual can only needs four peristaltic pumps. |
| 7.2 | Peristaltic pump | Intermittent type:Kolpalmer (Spirax Sarco) Watson-Marlow
Continous type:Kolpalmer (Spirax Sarco) Longerpump |
| 7.3 | The buyer’s confirmation: | |
| 8 | Inoculation and Seed transfer | |
| 8.1 | Inoculation | Inoculation approach:Flame inoculation Peristaltic pump + insert inoculation |
| 8.2 | subcultivation | Rigid tube seed transfer hose quick coupling seed transfer No |
| 8.3 | The buyer’s confirmation: | |
| 9 | Biological safety | |
| 9.1 | Exhaust disposition | High temperature sterilizer Chemical pool sterilization No |
| 9.2 | Sewage treatment | Dedicated pipe discharge mix emissions No |
| 9.3 | The buyer’s confirmation: | |
| 10 | Control function | |
| 10.1 | sterilization | Automatic sterilization of empty vessel or filled liquid vessel;Sterilization temperature: 100-130 ° C, sterilization time: 0-60 min;
Automatic sterilization of inlet and exhaust filters. Allowed tolerance of sterilization temperature:±0.1°C; Note: |
| 10.2 | Fermentation temperature | PID automatic control, plate heat exchanger for heating and cooling, cooling water for cooling, with circulating system;
Control range: cooling water temperature + 5-65 ° C, control precision: + / – 0.1 ° C, With real-time curve and historical curve, it can be tracked and controlled in sections. |
| 10.3 | pH | PID automatic control, peristaltic pump control, can choose one-way or two-way control;
Control range: 2-12ph, control accuracy: + 0.02 pH, resolution: 0.01pH It has a real time curve and a historical curve. |
| 10.4 | DO | PID automatic control, control air and O2 through mass flow meter;
It can be controlled in association with stirring speed and intake air flow, that is, the dissolved oxygen can be adjusted by changing the stirring speed and intake air flow; Control range: 0-100%, display range: 0-200%, resolution: 1% It has a real time curve and a historical curve. |
| 10.5 | Mixing speed | PID automatic control, stepless speed regulation control with governor;
Speed range: 20-200rpm; Control accuracy: plus or minus 0.5%, resolution: 1rpm It has a real time curve and a historical curve and can be tracked and controlled in section. |
| 10.6 | Tank pressure | PID automatic control, pneumatic control with proportional valve;
Control range 0-0.2 MPa, steady state precision: ±0.01MPa, resolution: 0.001 MPa It has a real time curve and a historical curve and can be tracked and controlled in section. |
| 10.7 | Ventilation flow | air 100L/min oxygen 3L/min nitrogen50L/min carbon dioxide 1L/min accuracy:±2%,resolution 0.1% |
| 10.8 | Material measurement | Yes None |
| 10.9 | Feeding control | Digital timing quantitative control, creep pump automatic replenishing; |
| 10.10 | antifoaming control | Yes None |
| 10.11 | Level control | Yes None |
| 10.12 | Exhaust gas sterilization | Yes None |
| 10.13 | Weighing system | Mettler weighing module |